    ASTM F2790-10(2014)

    Standard Practice for Static and Dynamic Characterization of Motion Preserving Lumbar Total Facet Prostheses
    • 발행일 : 2014-11-01
    • 발행기관 : ASTM
    즐겨찾기 장바구니 담기
    닫기

    상품이 장바구니에 담겼습니다.

    장바구니로 이동
    • [영어]PDF 131,600원

    • [영어]PRINT 131,600원

상세정보

분야 13.02 : 긴급 의료 서비스; 수색 및 구조
적용범위

1.1 This practice provides guidance for the static and dynamic testing of Lumbar Total Facet Prostheses (FP). These implants are intended to allow motion and lend support to one or more functional spinal unit(s) through replacement of the natural facets.

1.2 These test methods are intended to provide a basis for the mechanical comparison among past, present, and future non-biologic FP. These test methods allow comparison of devices with different methods of application to the lumbar spine. These test methods are intended to enable the user to mechanically compare devices and do not purport to provide performance standards for them.

1.3 These test methods describe static and dynamic tests by specifying load types and specific methods of applying these loads.

1.4 These test methods do not purport to address all clinically relevant failure modes for FP, some of which will be device specific. For example, these test methods do not address implant wear resistance under expected in vivo loads and motions. In addition, the biologic response to wear debris is not addressed in these test methods.

1.5 Requirements are established for measuring displacements and evaluating the stiffness of FP.

1.6 Some devices may not be testable in all test configurations.

1.7 The values stated in SI units are to be regarded as the standard with the exception of angular measurements, which may be reported in terms of either degrees or radians.

1.8 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.

국제분류(ICS)코드 11.040.40 : 외과. 보철술. 교정술용 이식
페이지수 8
Edition 10(2014)

이력정보

No. 표준번호 표준명 발행일 상태
1 ASTM F2790-10(2014)상세보기 Standard Practice for Static and Dynamic Characterization of Motion Preserving Lumbar Total Facet Prostheses 2014-11-01 구판
2 ASTM F2790-10(2019)e1상세보기 Standard Practice for Static and Dynamic Characterization of Motion Preserving Lumbar Total Facet Prostheses 2014-11-01 표준
3 ASTM F2790-10상세보기 Standard Practice for Static and Dynamic Characterization of Motion Preserving Lumbar Total Facet Prostheses 2010-01-01 구판

제공형태 닫기

  • PDF :

    직접 파일 다운로드 및 인쇄(마이페이지 확인)
  • 보안PDF :

    직접 파일 다운로드 및 인쇄(마이페이지 확인)* 단, 파일이동 및 복사 불가, 1회 다운로드 및 인쇄가능
  • PRINT :

    인쇄본 우편발송, 2~3일 소요(PDF파일 미제공)
  • BOOK :

    인쇄본 우편발송, 2~3일 소요(PDF파일 미제공)
  • ZIP :

    압축파일형태로 제공 (PDF, HTML, TXT, XLS 등으로 구성), 직접 파일 다운로드(마이페이지 확인)
  • CD/DVD :

    직접 파일 다운로드 및 인쇄(마이페이지 확인)
  • DB :

    별도 정보 제공
  • HARDCOPY :

    해외 배송 상품, 최대 3주 소요
  • 바인더 :

    바인더 우편발송, 2~3일 소요
  • 온라인구독 :

    WEB 접속 후 실시간 열람, 출력(1년간)